Jumper Medical Co.,Ltd.
|
Place of Origin: | Guangdong, China (Mainland) |
Add to My Favorites | |
HiSupplier Escrow |
Generator
|
|
Type
|
HF converter, DC High Voltage
|
Voltage
|
Adjustable in steps of 1 kV from 40 to 110 kV
|
mAs Values
|
Adjustable in 22 steps from 1 to 160 mAs, meeting R'10 seriers. It is 1, 1.3, 1.6, 2, 2.5, 3.2, 4, 5, 6.3, 8, 10, 12.5, 16, 20, 25, 32, 40, 50, 63, 80, 100, 125 and 160.
|
Output
|
3.5kW (100kV/35mA)
110kV 30mA
80kV 46mA
40kV 70mA
|
Working mode
|
Run in intermittent loading
|
Tolerance
|
kV +/- 10 %
mAs +/- (10 % + 0.2mAs)
|
X-ray Tube
|
|
Type
|
XD6A-0.5,3.5 / 125; Stationary Anode
|
Material of Target
|
Tungsten
|
Angle of Target
|
15°
|
Focal Spot Size
|
1.5mm
|
Normal Voltage
|
110kV
|
Max. Heat capacity
|
57kJ (76kHU)
|
X-ray Tube Assembly
|
|
Power
|
3.5kW
|
Max. Normal Voltage
|
110kV
|
Fixed filter
|
2.7mmAl
|
Isolation type
|
Oil
|
Temperature Protective
|
55°C Temperature relay
|
Max. Heat capacity
|
800kJ (1040kHU)
|
Max continuous heat dissipation
|
260W
|
Collimator
|
|
Max. Field at 1m
|
43cm X 43cm
|
Data for Radiation
|
|
Total Filter
|
2.8mmAl
|
Leakage Radiation
|
Avg. Air Kerma Rate less than1.0mGy/h at all places that around and 1m distance from focal spot.(Loading set 110kV and work 10000mAs max in 1 hour)
|
Irremovable filter :
|
2.8mmAl
|
Leakage Current
|
|
Leakage current to Ground
|
Normal:Not more than 2.5mA
Single Fault:Not more than 5mA
|
Leakage current to Covers
|
Normal:Not more than 0.1mA
Single Fault:Not more than 0.5mA
|
Power supply
|
|
Voltage
|
220V ac
|
Max. voltage tolerance
|
+/-10%
|
Frequency
|
50Hz
|
Max. Frequency deviation
|
+/-1Hz
|
Max. impedance
|
0.60W
|
Fusing
|
220V,16A Slow
|
Temperature
|
10 to 40°C
|
Relative Humidity
|
30 to 75%
|
Atmospheric Pressure
|
700 to 1060hPa
|
Movement and Weight
|
|
Distance from Tube Focal to Floor
|
50cm to 190cm, adjustable
|
Rotation around arm of Tube assembly
|
+90°to -90°
|
Rotation around axis of Tube assembly
|
-45°to +90°
|
Weight
|
102kg
|